The apparatus can be particles of oil, the cleanliness and pollution monitoring and analysis; hydraulic equipment and routine maintenance and repair; hydraulic components wear test; pure solvents and ultrapure water insoluble particulate testing.
The use of core technology - \\\\\\\"Laser measurement of particles\\\\\\\" and the classic method of using the oil industry NAS1638 and ISO4406, and in accordance with the requirements of users, built-users a wide range of standards required.
Quoted precision piston pump and ultra-precision electromagnetic flow control system, to achieve a constant injection speed and injection volume precision dual control, the sampling volume of 1ml ~ infinity arbitrarily set, the accuracy.
Sensor using the classic \\\\\\\"laser particle measurement,\\\\\\\" a dedicated sensor, more suited to NAS1638 and ISO4406.
Built-in threshold and pulse resistance curve particle size, particle size can be set channel value.
Integrated automatic sampling positions, equipped with pressure measuring system, can be positive / negative pressure, so that equipment can be degassed samples and the detection of high-viscosity samples.
The use of large-screen liquid crystal display, touch screen menu operation, keyboard, touch-dual-input function and the whole aesthetic appearance.
Feature-rich data-processing; can be given in accordance with standard oil grades such as mapping the distribution histogram.
Built-in operating system and micro-printer, no external computer and the printer and print directly tested.
With a standard RS232 serial port can be an external computer storage test results to facilitate data classification, retrieval.
Instrument parameters:
Viscosity measurement: 0 ~ 350M2 / S;
Measuring range: 1¦Ìm ~ 450¦Ìm;
Particle concentration: ¡Ý 15000 tablets / ml;
Sensitivity: 1¦Ìm (ISO 4402) or 4¦Ìm (C) (ISO 11171, GB/T18854);
Measurement channels: 8 channels (which can be arbitrarily set);
Injection volume precision: better than ¡À 1%;
Test time: 12.5s, can be arbitrary;
Injection speed: 5mL/min ~ 60mL/min;
Sample temperature: room temperature ~ 65 ¡æ
Pressure range: vacuum1.0 ~ 1.0Pa;
Resolution: <10% (ISO 11171);
Coincidence error limit: 10,000 / mL (5% coincidence error);
Data Output: built-in printer;
Standard Interface: RS232 interface which can be accessed by computer;
Power Supply: 100 ~ 245V, 50Hz ¡À 1%, <80W.
 


 Through working together with our customers to understand their operational and organizational needs, our services and solutions are customized to provide the most cost effective through-life support. We liaise with the customer to create the right solution incorporating the appropriate capabilities of GE. These can include inventory management and supply, spares, repairs, modifications, reliability enhancements, capability upgrades, data management and information exploitation. All our flexible services and solutions are backed up by 24/7 technical and logistics support, and the unique knowledge, expertise and know-how of our GE Original Equipment Manufacturers (OEMs).!